| CERTIFICATIONS | OUR Bomb Calorimeter ARE CE Certified Chambers . We as company is ISO 9001:2008 Certified Company. |
| MODEL & FEATURES | HAMCO 6F Automatic Model With
Single Unit. No assembly Required. Chiller external with 2 mtr tubing. Color Display with PLC Touch screen with 1000 Data Storage Memory. Sample Preparation Manual. Analysis time not more than 10 Min If in predict mode analysis time less then 8 min Oxygen Filling Automatic. Temp Rise Detection Automatic. Calculation Automatic. Correction fuse wire,Acid correction, spike correction, sulfur correction. Bucket Fixed.. Bomb removable. Water filling automatic : Water Handling system with auto water filling and draining. Bomb wash manual. Jacket controlled with accuracy of +/- 0.05 deg C. Printer & Reporting : YES. Standardization Automatic with 10 result Memory. You can Do average or Manual Entry for Standardization Value Or you can calculate. Reporting Units Cal/gm,BTU/Lb , Kcal/Kg , or selectable. Give You the Final Report where you add Ash content, Moisture, Sulphor content, Etc |
| Highly accurate results with precision 0f 0.1% RSD. | |
| TEST METHOD | Determines the Gross Calorfic Value of any solid or liquid sample by determining the water equivalent by burning 1 gm of Known CV sample. In isoperibol mehod the jaket will be at controlled temperature to isolate buket
from environment and same water will be used as bucket water. |
| TEMP RANGE/ RESOLUTION | Temperature Range 15C to 35C ,
Bucket Temperature resolution +/- 0.0001C . Jacket temperature resolution 0.01C. |
| METHOD | ISOPERIBOL |
| STANDARD | Confirm to the specification ASTM D 4809(COAL,COKE), D 240(LIQUID HYDROCARBON), ASTM D 5865(LDO/FO) , D E711 , IS -1350- 1966, BS
1016 PART 5-1967 , IP 12/63 T. |
|
CONSTRUCTION STANDARD MODEL |
Outer body made of GI Powder coated.
Inner jacket made of SS 304 with Insulation. Inner Jacket Wrapped with high insulation cloth to minimize the Heat correction factor to boost the accuracy. Bomb bucket is made of copper with Plating as per ASTM Standard to get the best of accuracy. Bomb Support to Ensure Correct Positioning of Bomb in Bucket. Compact chiller to maintain jacket temperature and |
|
SS BOMB |
Made of SS 316
Tested hydraulically for pressure upto 300 Kg/Cm2 Easy to Handle & Operate Easy Maintenance. Only Periodically O ring (Usually after 500 Test) or Valves(Usually around 1000 Test) need to be changed. Control panel is provided with Fuse protection in case of electrode short or Electrode with body short. Hamco will Provide O ring set , Schrader Valve & Fuse set one extra as spares. |
|
SAMPLE PREPARATION |
Sample Preparation need to do manual.
If any Non Evaporatable liquid sample is there , we need to pour 1 gm of sample in crucible. If any Evaporatable liquid sample is there , we need to weigh it , put it in gelatin capsul and put the capsule in the Crucible. If any Powder form solid sample is there , we need to make 1 gm tablet and weigh it and put it in crucible. If any fiborous solid material is there, we need to crush it, Pass it through sieve 20 and than make 1 gm tablet , weigh it and put it into the crucible. |
| OXYGEN FILLING | Oxygen Filling Will be Auto.
It is Inbuilt. The oxygen filling is very minute that if we use powder form instead of making pellet , the sample will not be spilled in the bomb. |
|
TEMP. RISE DETECTION |
Temp. Rise Detection Automatic.
Stirrer Start Auto. Firing Auto Stabilization mode : Auto No Need to do anything , Just Press one button and you are free for the test. You can Observe Time Vs Temp Reading On Monitor Screen. |
|
CALCULATION & REPORTING |
Calculation Automatic.
For CV Determination, We just need to Mass of the sample Manually and press start Button , thats all. It also has all the interlock inbuilt. Like after standarzation expiry , it gives you Pop Up. If you forget to connect stirrer wire or Fuse wire or Bomb wire, Ignition wire electrode short , Charging Fail etc. |
|
JACKET TEMPERATURE |
Jacket temperature will be controlled by external chiller.
Jacket temperature resolution is 0.01 C. Jacket temperature accuracy +/- 0.05 C. Highly accurate jacket controlling. |
|
WATER FILLING AUTOMATIC |
Controlled jacket temperature once get stabilized, you will get calorimeter ready message.
Once jacket is ready and once we start the test bucket will be filled automatically and once analysis done will back to chiller for cooling. |
|
OHER FEATURES |
SS Bomb Removal type.
Bucket Fixed type. Chiller external. Water filling automatic. |
| POST ANALYSIS | Below analysis can be done once CV determination done. This will be done manual and will be feeded in the Machine to get combined report. Accessories required to do titration and drying will be at customer scope.
Ash content. Moisture. Sulfur determination. Fluoride. |
| CV DETERMINATION RANGE & ACCURACY | CV Range 1000 to 20000 Cal /gm.
RSD +/- 0.1%. In CRM mode the RSD +/- 0.0.5%. |
| EXPERIMENT DURATION | Less than 10 Min
In predict mode analysis less then 8 min |
| Desktop based Utility software with Live screen for Temp Monitoring Graph and Pop up on PC regarding Start of Test, |
| SOFTWARE | Mode of the Test, Result of the test.
It has User password login with Multiple user level Access. It can print the data on Attractive Report format with Company Logo and detail. You can Filter the data on Sample Name wise, Operator Name wise, Supplier Name wise, Date & Time wise & Batch Wise. |
|
CONTROL PANEL |
PLC Touch Screen Color Display.
IP 22 Protection. You Can Do CV Determination Test, Standardisation Test , Manual Test, Instrument Test, Oxygen Filling Test and Can scroll Down 400 Data History. Select Reporting Units. Mains Fuse For Load Protection. Ignitor Fuse For Electrode short circuit protection. Interlock for Ignition wire wrong connection. Interlock for stirrer connection cable if loosen. 12V DC power Supply to Ignition Circuit. LED to show standby Status, Program running status, Program Over status. |
|
APPLICATION |
For CV Determination of Any Solid or Liquid.
This Semiautomatic Type of Bomb calotrimeter , This we suggest to industries , colleges , institutes , Lab etc. Still you want advance model we have the automatic one too. |
| HEATING LOAD | Max Up to 1 KW. |
| POWER SUPPLY | 230 V Ac , 5 Amp , 50/60 Hz Power Supply |
|
ACCESSORIES SUPPLIED ALONG WITH INSTRUMENTS |
S.S. Bomb (CAT No : HAMCO 6AC 1)
Main Unit. Connecting Tubes (4mm High Pressure tube with Swag Lock) for filling Oxygen in the Bomb: 20 Mtr Spanner for Oxygen valve & copper tube Connection. Oxygen Control Valve. (CAT No : HAMCO 6AC 2) Pellet Press. (CAT No : HAMCO 6AC 3) Connecting Wire [Electrical]. Hook for Lifting Bomb. (CAT No : HAMCO 6AC 4) Crucible. [Stainless Steel] (CAT No : HAMCO 6AC 5) Cotton Reel(CAT No : HAMCO 6AC 6) :01 NO. (Approximately 200 Meters) Bomb O‟ Ring(CAT No : HAMCO 6AC 7) : 01 No. Teflon tape : 01 No. Valve Body O‟ Ring (CAT No : HAMCO 6AC 8) : 01 No. Ignition Wire (CAT No : HAMCO 6AC 9) : 100 Meters. Schrader Valve (CAT No : HAMCO 6AC 10) : 01 Nos. Bursting Discs for Safety Device 12 Nos. : 1 set. (CAT No :HAMCO 6AC 11) Benzoic Acid 100 Tablets of 1 Gm of Known CV Certified Imported . { 6319 Cal / gram +/- 20 Cal/gm} Valve Key. {4 sided key used for replacing/tightening Schrader valve, See (p) in precautions on page 12}. Chiller with tubing. 220 micron test sieve. Operating manual. |
|
DOCUMENTS SUPPLIED ALONG WITH THE INSTRUMENTS. |
IQ,OQ,PQ OF COMPLETE APPARATUS
INSTRUCTION MANUAL HMI SOP SS BOMB PRESSURE TEST CERTIFICATE. HMI TEMPERATURE INDICATOR CERTIFICATE. PRESSURE GAUGE TEST CERTIFICATE. SS BOMB MATERIAL TEST CERTIFICATE. BENZOIC ACID CV STANDARD TEST CERTIFICATE. HYDRAULIC PRESSURE TEST CERTIFICATE. ASTM METHOD COMPLIANCE CERTIFICATE. HIGH PRESSURE NYLON TUBING PRESSURE WITHSTANDING TEST CERTIFICATE. |
|
UTILITIES (PREINSTALLATION REQUISITE) |
Oxygen Cylinder Industrial Type with 99% Purity with Key.
Weighing Balance of reputed Make with 0.01 mg. Distilled Water Approx 15 Ltr. PC for software. |
| TYPE OF BOMB CALORIMETER | FULLY AUTOMATIC ISOPERIBOL BOMB CALORIMETER SINGLE UNIT WITH PLC TOUCH SCREEN COLOR DISPLAY WITH CHILLER WITH
AUTO WATER HANDLING SYSTEM |