FULLY AUTOMATIC ISOPERIBOL BOMB CALORIMETER SINGLE UNIT (NO ASSEMBLY REQUIRED) WITH TOUCH SCREEN PLC AS PER ASTM D 4809/5865 HAMCO 6G WITH EXTERNAL CHILLER AND AUTO WATER HANDLING SYSTEM

| CERTIFICATIONS | OUR Bomb Calorimeter ARE CE Certified Chambers. We as company is ISO 9001:2008 Certified Company. |
| MODEL & FEATURES | HAMCO 6G Automatic Model With
Single Unit. No assembly Required. Chiller external with 2-meter tubing. Color Display with PLC Touch screen with 1000 Data Storage Memory. Sample Preparation Manual. Analysis time approximately 10 Min per sample Oxygen Filling Automatic with auto shut off. Temp Rise Detection Automatic. Calculation Automatic. Correction fuse wire, Acid correction, Moisture correction, spike correction, ash, nitrogen, sulfur & hydrogen correction. Bucket Fixed. Bomb removable. Water filling automatic: Water Handling system with auto water filling and draining. Bomb wash manual. Jacket controlled with accuracy of +/- 0.05 deg C. Printer built-in & Reporting: YES. Standardization Automatic with 10 result Memory. You can Do average or Manual Entry for Standardization Value Or you can calculate. Reporting Units Cal/gm, BTU/Lb., Kcal/Kg, or selectable. Give You the Final Report where you add Ash content, Moisture, Sulphur content, Etc. Highly accurate results with precision 0f 0.1% RSD. |
| Calibration facility Auto / manual using certified standard (benzoic acid).
Equipment measuring resolution -0.0001 Cal/gm. or better. Operator time less than one minute. Operating Temp: 30oC or more Direct analysis of powdered sample / pellet sample/ Liquid sample. Sample size – 0.5 to 1.5 gm |
|
| TEST METHOD | Determines the Gross Calorific Value of any solid or liquid sample by determining the water equivalent by burning 1 gm of Known CV sample. In isoperibol method the jacket will be at controlled temperature to isolate bucket from environment and same water will be used as bucket
water. |
| TEMP RANGE/ RESOLUTION | Temperature Range: 15C to 35C,
Bucket Temperature resolution +/- 0.0001C. Jacket temperature resolution 0.01C. |
| METHOD | ISOPERIBOL |
| STANDARD | Confirm to the specification ASTM D 4809(COAL, COKE), D 240(LIQUID
HYDROCARBON), ASTM D 5865(LDO/FO), D E711, ISO 1928, IS -1350- 1966, BIS 1350 (PART 2)-2022, BS 1016 PART 5-1967, IP 12/63 T. |
|
CONSTRUCTION STANDARD MODEL |
Outer body made of GI Powder coated.
Inner jacket made of SS 304 with Insulation. Inner Jacket Wrapped with high insulation cloth to minimize the Heat correction factor to boost the accuracy. Bomb bucket is made of copper with Plating as per ASTM Standard to get the best of accuracy. Bomb Support to Ensure Correct Positioning of Bomb in Bucket. Compact chiller to maintain jacket temperature and |
|
SS BOMB |
Made of SS 316
Conforms to ASTM E144 / IS350 (part-II):2022 Tested hydraulically for pressure up to 300 Kg/Cm2 5000 guaranteed number of ignitions before re-hydraulic test of Bomb/vessel. Easy to Handle & Operate Easy Maintenance. Only Periodically O ring (Usually after 500 Test) or Valves (Usually around 1000 Test) need to be changed. Control panel is provided with Fuse protection in case of electrode short or Electrode with body short. Hamco will Provide O ring set, Schrader Valve & Fuse set one extra as spares. |
|
SAMPLE PREPARATION |
Sample Preparation need to do manual.
If any non-evaporable liquid sample is there, we need to pour 1 gm of sample in crucible. If any Evaporable liquid sample is there, we need to weigh it, put it in gelatin capsule and put the capsule in the Crucible. If any Powder form solid sample is there, we need to make 1 gm tablet and weigh it and put it in crucible. If any fibrous solid material is there, we need to crush it, pass it through sieve 20 and then make 1 gm tablet, weigh it and put it into the crucible. |
|
OXYGEN FILLING |
Oxygen filling will be automatic with auto shut off.
It is Inbuilt. The oxygen filling is very minute that if we use powder form instead of making pellet, the sample will not be spilled in the bomb. |
|
TEMP. RISE DETECTION |
Temp. Rise Detection Automatic.
Stirrer Start Auto. Firing Auto Stabilization mode: Auto No Need to do anything, Just Press one button and you are free for the test. You can Observe Time Vs Temp Reading on Monitor Screen. |
|
CALCULATION & REPORTING |
Calculation Automatic.
For CV Determination, we just need to Mass of the sample Manually and press start Button, thats all. It also has all the interlock inbuilt. Like after standardization expiry, it gives you Pop Up. If you forget to connect stirrer wire or Fuse wire or Bomb wire, Ignition wire electrode short, Charging Fail etc. |
|
JACKET TEMPERATURE |
Jacket temperature will be controlled by external chiller.
Jacket temperature resolution is 0.01 C. Jacket temperature accuracy +/- 0.05 C. Highly accurate jacket controlling. |
|
WATER FILLING AUTOMATIC |
Controlled jacket temperature once gets stabilized you will get calorimeter ready message.
Once jacket is ready and once we start the test bucket will be filled automatically and once analysis done will back to chiller for cooling. External water jacket with temperature controller of precision as per ASTM Bucket volume should be minimum 2000 ml. |
|
OHER FEATURES |
SS Bomb Removal type.
Bucket Fixed type. Chiller external. Water filling automatic. |
|
POST ANALYSIS |
Below analysis can be done once CV determination done. This will be done manual and will be feeded in the Machine to get combined report.
Accessories required to do titration and drying will be at customer scope. Ash content. Moisture. Sulfur determination. Fluoride. |
| CV DETERMINATION RANGE &
ACCURACY |
CV Range 1000 to 20000 Cal /gm.
RSD +/- 0.1%. |
| ANALYSIS
DURATION |
Approximately 10 Min or less. |
|
SOFTWARE |
Desktop based Utility software with real time service diagnostics including ambient chart of the instrument, temperature, pressure, and communication diagnostics. Live screen for Temp Monitoring Graph and Pop up on PC regarding Start of Test, Mode of the Test, Result of the test.
Software will have 10-year license with provision for upgradation. It has User password login with Multiple user level Access. It can print the data on Attractive Report format with Company Logo and detail. You can Filter the data on Sample Name wise, Operator Name wise, Supplier Name wise, Date & Time wise & Batch Wise. |
|
CONTROL PANEL |
PLC Touch Screen Color Display.
IP 22 Protection. You Can Do CV Determination Test, Standardization Test, Manual Test, Instrument Test, Oxygen Filling Test and Can scroll Down 400 Data History. Select Reporting Units. Mains Fuse for Load Protection. Ignitor Fuse for Electrode short circuit protection. Interlock for Ignition wire wrong connection. Interlock for stirrer connection cable if loosen. 12V DC power Supply to Ignition Circuit. LED to show standby Status, Program running status, Program Over status. Inbuilt data storage of along with the option to transfer data to external storage device. |
| APPLICATION | For CV Determination of Any Solid or Liquid. |
| HEATING LOAD | Max Up to 1 KW. |
| POWER SUPPLY | 230 V Ac, 5 Amp, 50/60 Hz Power Supply |
|
ACCESSORIES SUPPLIED ALONG WITH INSTRUMENTS |
S.S. Bomb (CAT No: HAMCO 6AC 1)
Main Unit. Connecting Tubes (4mm High Pressure tube with Swag Lock) for filling Oxygen in the Bomb: 20 Meter Spanner for Oxygen valve & copper tube Connection. Oxygen Control Valve. (CAT No: HAMCO 6AC 2) Pellet Press. (CAT No: HAMCO 6AC 3) Connecting Wire [Electrical]. Hook for Lifting Bomb. (CAT No: HAMCO 6AC 4) Crucible. [Stainless Steel] (CAT No: HAMCO 6AC 5) Cotton Reel (CAT No: HAMCO 6AC 6):01 NO. (Approximately 200 Meters) Bomb O Ring (CAT No: HAMCO 6AC 7): 01 No. Teflon tape: 01 No. Valve Body O Ring (CAT No: HAMCO 6AC 8): 01 No. Ignition Wire (CAT No: HAMCO 6AC 9): 100 Meters. Schrader Valve (CAT No: HAMCO 6AC 10): 01 Nos.
Bursting Discs for Safety Device 12 Nos.: 1 set. (CAT No: HAMCO 6AC 11) Benzoic Acid 100 Tablets of 1 Gm of Known CV Certified Imported. [6319 Cal / gram +/- 20 Cal/gm] Valve Key. (4 sided key used for replacing/tightening Schrader valve, See (p) in precautions on page 12). Chiller with tubing. 220-micron test sieve. Operating manual. Additional accessories as mentioned in Technical Specifications. |
|
DOCUMENTS SUPPLIED ALONG WITH THE INSTRUMENTS. |
IQ, OQ, PQ OF COMPLETE APPARATUS
INSTRUCTION MANUAL HMI SOP SS BOMB PRESSURE TEST CERTIFICATE. HMI TEMPERATURE INDICATOR CERTIFICATE. PRESSURE GAUGE TEST CERTIFICATE. SS BOMB MATERIAL TEST CERTIFICATE. BENZOIC ACID CV STANDARD TEST CERTIFICATE. HYDRAULIC PRESSURE TEST CERTIFICATE. ASTM METHOD COMPLIANCE CERTIFICATE. HIGH PRESSURE NYLON TUBING PRESSURE WITHSTANDING TEST CERTIFICATE. |
|
UTILITIES (PREINSTALLATION REQUISITE) |
Oxygen Cylinder Industrial Type with 99% Purity with Key.
Weighing Balance of reputed Make with 0.01 mg. Distilled Water Approx 15 L. PC for software with every upgradation if any. |
|
|
TYPE OF BOMB CALORIMETER |
FULLY AUTOMATIC ISOPERIBOL BOMB CALORIMETER SINGLE UNIT WITH PLC TOUCH SCREEN COLOR DISPLAY WITH CHILLER & AUTO WATER HANDLING SYSTEM |
|
|
Display/ Interface |
Main system will be able to work stand alone. In addition to, it should have direct interface to connect with printer and balance through RS 232 / USB port. | |
|
SAFETY LOCK SYSTEM |
Safety Lock System for Bomb for users safety. The Calorimeter Lid/Head should automatically Lock electronically when the Instrument is in operation mode. Safety of Bomb should be assured as per ASTM E144 / IS350 (part-II):2022 or equivalent. |
|
| SELF DIAGNOSTIC SYSTEM | Instrument must have facility for Extensive Diagnostic Function for problem investigation or any other fault arising in the
instrument. |
|
| MODEL NO | HAMCO 6G | |
| WARRANTY | One year warranty against manufacturing defects. | |