Automatic Isoperibol Bomb Calorimeter

HAMCO6F Automatic Isoperibol Bomb Calorimeter With PLC With Printer As Per Astm D 5865

WITH EXTERNAL CHILLER AND 2000ML AUTO BURRETTE

Automatic

Test method: – determines the gross Calorfic value of any solid or liquid sample by determining the water equivalent by burning 1 gm of known CV sample.

Standard: – confirm to the specification ASTM D 5865, D 240, DE711, IS -1350- 1966, BS 1016 PART 5-1967, IP 12/63 T.

Features & benefits: –

  • Type of method: – isoperibol
  • Method of detecting temp. Rise : automatic
  • Method of detecting calorific value : automatic.
  • With External chiller with 2 Mtr Tubing.
  • Bucket type : Removable.
  • Bomb type : Removable.
  • Water filling semi automatic : Auto 2000 ml burette can feed the metered water at Controlled jacket temperature.
  • Jacket Resolution : 0.01 Deg C
  • Jacket Accuracy : 0.05 Deg C.
  • Accuracy :- 0.1% RSD
  • With 1000 sample consumable kit.
  • Touch Screen PLC.
  • Least Count is 0.0001 Deg C & Accuracy +/- 1% RSD in Benzoic Acid.
  • Standard packet of 100 Tablet of Benzoic acid supplied along with the instrument with certificate.
  • Automatic measurement and calculation of Calorfic value / water equivalent.
  • 3 Color Screen TFT.
  • Single Unit , No Individual accessory. Just Plug in start , No assembly required.
  • Thermal printer interface with printout report.
  • Auto alarm on completion of test.
  • Pc interface for data recording.
  • Final report can enter the value of Ash content, Moisture % , Sulphor %
  • Weight of the sample to be filled by keyboard.
  • Calorimeter vessel made of Copper Electroplated
  • Bomb support to ensure correct positioning of bomb in vessel.
  • Ambient working temp range : 15 to 35 degree c
  • Heat capacity :- 2200 to 2500 cal/deg c
  • Process time: – 10 to 12 minute.
  • CV Range :- 500 to 20000 Cal/Gm.
  • GCV Resolution :- 0.1 Cal/gm
  • 1000 Data storage Memory.
  • Reporting Units can be selected cal/gm, Kcal/Kg, BTU/LB, MJ/Kg or others.
  • Standardization EE value can be stored of last 10 reading. The value can be averaged, erased or selected. We can enter Fixed EE value too.
  • Thermoelectric fuse correction, acid correction, sulphor correction value can be entered and automatically corrected in Final value.
  • Temperature resolution :- 0.0001 deg c
  • Oxygen bomb endure pressure : 300 kg/cm
  • Humidity requirement : less than 85%
  • Power supply: 230a VC, 50 Hz, 10 amp.
  • Heating load: 2 kW.
  • Control panel/display :-
    1. Digital display of temperature.
    2. Auto temp. Rise detection based on microprocessor chip based. User has to press one button and is free for doing other thing while microprocessor will take reading automatically.
    3. Automatic reading for temp. Rise calculation.
    4. All the necessary interlock for stirrer, ignition wire continuity, wire loose, over load etc.
    5. Electrode short indication.
    6. Continuity test indication.
    7. Auto stop of the stirrer after completion of test.
    8. Buzzer horn after completion of the test.
  • Safety interlocks/ relay :-
    1. Over load protection.
    2. Electrode short protection & indication.
    3. 1-phase indication lamp.
    4. Emergency switch off
  • Construction :-
    1. Double walled construction, thermally insulated by pouring approx. 10 ltr water in the outer jacket.
    2. Made of full stainless steel.

Accessories will be supplied with equipment: –

  • S. Bomb (CAT No : HAMCO 6AC 1)
  • Main
  • Connecting Tubes (4mm High Pressure tube with Swag Lock) for filling Oxygen in the Bomb: 20 Mtr
  • Spanner for Oxygen valve & copper tube
  • Oxygen Control Valve. (CAT No : HAMCO 6AC 2) Pellet Press. (CAT No : HAMCO 6AC 3)
  • Connecting Wire [Electrical].
  • Hook for Lifting Bomb. (CAT No : HAMCO 6AC 4)
  • [Stainless Steel] (CAT No : HAMCO 6AC 5)
  • Cotton Reel(CAT No : HAMCO 6AC 6) :01 NO. (Approximately 200 Meters)
  • Bomb O‟ Ring(CAT No : HAMCO 6AC 7) : 01
  • Teflon tape : 01
  • Valve Body O‟ Ring (CAT No : HAMCO 6AC 8) : 01 No.
  • Ignition Wire (CAT No : HAMCO 6AC 9) : 100
  • Schrader Valve (CAT No : HAMCO 6AC 10) : 01
  • Bursting Discs for Safety Device 12 Nos. : 1 set. (CAT No : HAMCO 6AC 11)
  • Benzoic Acid 100 Tablets of 1 Gm of Known CV Certified Imported .

{ 6319 Cal / gram +/- 20 Cal/gm}

  • Valve {4 sided key used for replacing/tightening Schrader valve, See (p) in precautions on page 12}.
  • Chiller with tubing.
  • 2000 Ml burette with burette stand
  • 220 micron test sieve.
  • Operating manual.

Spares required for 5 year trouble free operation: –

  • Cotton reel (cat no: hamco 6ac 6):05 no. (approximately 200 meters)
  • Bomb o ring (cat no: hamco 6ac 7): 02 no.
  • Valve body o ring (cat no: hamco 6ac 8): 02 no.
  • Ignition wire (cat no: hamco 6ac 9): 100 meters.
  • Schrader valve (cat no: hamco 6ac 10): 01 nos.
  • Burrette 2000 Ml : 2 Nos.
  • SS Bomb : 01 No.

Electrical requirement

  • 220-240v, 50/60hz, single phase, 10.9a
  • Or 115v, 50/60hz, single phase, 20.8a

Dimension: – ixwxh,in.(cm)

16×21.5×14.5 (41x55x37)
net weight: 64 lbs (29kg)

21 X 14.5 X 16 (53x37x40)

net weight :77 lbs(35kg)

Shipping information: –

Shipping weight: 174 lbs (79kg)
dimensions: 20 cu. Ft.

Ordering information: –

Catalogue no Product description Electrical requirement
Hamco 6F Automatic Isoperibol Bomb Calorimeter 220-240v, 50/60hz
Hamco 6F exp Automatic Isoperibol Bomb Calorimeter 115v, 50/60hz
Hamco 6ac1 SS bomb
Hamco 6ac2 Oxygen control valve

 

 

 

Hamco 6ac3 Pellet press

 

 

 

Hamco 6ac4 Hook for lifting bomb

 

 

 

Hamco 6ac5 Crucible
Hamco 6ac6

 

Cotton reel
Hamco 6ac7

 

Bomb o ring
Hamco 6ac8

 

Valve body o ring
Hamco 6ac9

 

Ignition wire
Hamco 6ac10

 

Schrader valve
Hamco 6ac11

 

Bursting disc

 

  • What is ISOPERBOL Bomb calorimeter.

The term “isoperibol” means constant pressure. In an isoperibol calorimeter, the bomb is submerged in a known volume of water, and the whole system is enclosed in an insulated container. The water acts as a heat sink, absorbing the heat released during the combustion reaction. Insulated container which normally contain highly precision temperature water which act as insulator from the surrounding and this constant temperature water makes the system more accurate.

  1. Benefits of using Hamco 6F( Automatic Isoperibol Bomb Calorimeter) over Hamco 6E
  2. Much More accurate results :- As in Isoperibol method due to high precision of the jacket water and same water will be used as bucket makes it much more accurate and isolate from room temperature fluctuation.
  3. Easier Calibration :- Isoperibol calorimeters are generally easier to calibrate compared to isothermal calorimeters. The water jacket surrounding the bomb helps in stabilizing the temperature and pressure conditions, simplifying the calibration process and ensuring more consistent results. The duration of calibration is much higher compared to isothermal bomb calorimeter. Normally frequency to calibrate isoperibol bomb calorimeter is around 3 to 4 months.
  4. Reduced Heat Loss: The insulated design of the isoperibol calorimeter minimizes heat loss to the surroundings during the combustion process. This insulation helps maintain the accuracy of temperature measurements and reduces errors that could arise from external temperature fluctuations.
  5. More Accurate Heat Measurements: By maintaining constant pressure conditions and minimizing heat loss, isoperibol calorimeters generally provide more accurate measurements of the heat of combustion compared to isothermal calorimeters. This accuracy is crucial in research and industrial applications where precise energy content determination is essential.
  6. Why HAMCO Make Automatic Isoperibol Bomb calorimeter ?
  • Hamco has more than decade experience in manufacturing of automatic isoperibol bomb calorimeter.
  • Hamco has successfully delivered and commissioned more than 1000 automatic Bomb calorimeter in India and across the globe.
  • Hamco is constantly Innovating and bring best of best technology and feature first in the market which allows our customers to win.
  • Hamco is one of the leading manufacturer of Automatic isoperibol bomb calorimeter in india.
  • Hamco is Manufacturer of Precise Automatic bomb calorimeter manufacturer in india.
BOMB CALORIMETER DETAILS
SR.NO PARAMETER M/s Hindustan apparatus, Mumbai , Model HAMCO-6E M/s Hindustan apparatus, Mumbai , Model HAMCO-6F
1 Type of bomb Calorimeter FULLY AUTOMATIC PLC BASED FULLY AUTOMATIC PLC BASED WITH EXTERNAL CHILLER AND SEMIAUTO WATER FILLING
2 METHOD AUTOMATIC ISOTHERMAL BOMB CALORIMETER AUTOMATIC ISOPERIBOL BOMB CALORIMETER
3 Measuring principal and method ISOTHERMAL AND ASTM D4893,240,5865 ISOPERIBOL AND ASTM D4893240,5865
4 GCV Capacity ( range) 1000 TO 20000 KILO CALORY P/KGM 500 TO 20000 KILO CALORY P/KGM
5 Testing precision( precision mode) 1% RSD ON BENZOIC ACID STANDARD SAMPLE 0.1% RSD ON BENZOIC ACID STANDARD SAMPLE
6 Testing precision(predictive mode) 1% RSD 0.1% RSD
7 Resolution of GCV 1 CALORIE PER GRAM 0.1 CALORIE PER GRAM
8 Nos of Analysis per Hour MAXIMUM 4 MAXIMUM 4 TO 6
9 Analysis Time 10 TO 15 MIN 10 TO 12 MIN
10 Temp. Measurement Accuracy 0.0001 Deg C 0.0001 Deg C
11 Temp. Measurement Resolution 0.0001 Deg C 0.0001 Deg C
12 Permisible Value Ambient Temp. 15 to 40 Deg C 15 to 40 Deg C
13 Gas (Oxygen)Filling AUTO AUTOMATIC
14 Bomb washing MANUAL MANUAL
15 Bucket filling MANUAL SEMIAUTOMATIC WITH MEASURE BURETTE
16 Wire tying MANUAL MANUAL
17 Bomb pressure release AUTOMATIC AUTOMATIC
18 Water circulation in outer jacket PUFF INSULATED WATER JACKETING AUTOMATIC WITH EXTERNAL CHILLER WITH ACCURACY OF +/- 0.05 DEG C
19 Digital Weighing Scale For Auto Weight Transfer by Softwire MANUAL ENTRY SUITABLE TO MATLER OR MANUAL ENTRY
20 Software WEB BASED SOFTWARE WITH ONLINE AND OFFLINE MONITORY WEB BASED SOFTWARE WITH ONLINE AND OFFLINE MONITORY
21 Report Customization NOT POSSIBLE POSSIBLE WITH SOFTWARE ONLY
22 Computer Requirement YES AT BUYSER SCOPE YES AT BUYSER SCOPE
23 Installation & commissioning YES REQUIRED YES REQUIRED
24 Accessories 500 Samples 2000 Samples
25 Power consumption per hrs use 2 UNITS PER DAY MAXIMUM 2 UNITS PER DAY MAXIMUM
26 Room Temperature NEED AC OR CONTROLLED ROOM TEMP NOT REQUIRED
27 Chiller NOT INCLUDED INCLUDED
28 Coal sample preparation Kit NOT INCLUDED INCLUDED

 

Get In Touch