NewPDualBOD

 

 

CERTIFICATIONS OUR CHAMBERS ARE CE Certified Chambers . We as company is ISO 9001:2008 Certified Company.
FEATURES Consist of Dual Chamber (Hamco 71 series) + Ethernet Based Connectivity + HMI PLC ( 7 Touch Screen PLC Display ) + DS ( Door Access System ) + SOFTW-21 (Web browser based Software Part 11 21 CFR + SCN-4 ( 4 Temp sensor ) +EX.HTR (External Hooter for alarm

Transmission) + GSM (Mobile SMS Alert) +EMAIL ( Email alert)+ VAL ( Validation)

RANGE/ ACCURACY / UNIFORMITY UPPER CHAMBER:

Temperature Range Ambient+5C to 60C , Temperature Accuracy +/- 0.5C , Temperature Uniformity : +/- 1C .

LOWER CHAMBER:

Temperature Range Ambient+5C to 60C , Temperature Accuracy +/- 0.5C , Temperature Uniformity : +/- 1C .

 

CONSTRUCTION GMP MODEL

Double walled , gap between walls filled with PUF Insulation. Inner Body made of SS 304 Mirror Finished, Outside Body made of SS 304 Matt Finished . Double door : Inner full Viewing door and outer door insulated withlock and key arrangement. 50mm Validation Port with silicon rubber seal to insert sensorforvalidationpurpose. 3 Thick CFCFree Puff Insulation. Stainless steel Trays. Adjustable Tray Height adjustment.

Unit Mounted on castor wheels with front lock arrangement. Stainless steel Rod Trays.

Adjustable Tray Height adjustment.

LIGHTING Floroscent Light are provided , which automatically On when we open the door.
 

CHAMBER PRINCIPLE

UPPER CHAMBER:

Air circulation by Circulation Motor & Fan will continuously sucks the air from the chamber and Passes it through Air heater, Evaporator and through it back tochamber and same air will be recirculated to achieve desired Temperature stability inside chamber. Our design reduces the power consumption by 50% .

LOWER CHAMBER:

Air circulation by Circulation Motor & Fan will continuously sucks the air from the chamber and Passes it through Air heater, Evaporator and through it back tochamber and same air will be recirculated to achieve desired Temperature stability inside

chamber. Our design reduces the power consumption by 50% .

 

 

 

 

 

 

ICH GUIDELINES

UPPER CHAMBER:

Fully Suitable to carry out Stability Studies For Conditions Such As :

32.5C

37.5C

42.5C

47.5C

52.5C

57.5C

Or any other conditions within range.

The underline temp maintain only if room temp is 25C.

LOWER CHAMBER:

Fully Suitable to carry out Stability Studies For Conditions Such As :

32.5C

37.5C

42.5C

47.5C

52.5C

57.5C

Or any other conditions within range.

The underline temp maintain only if room temp is 25C.

Other Than this cycle Please Mention it before order.

APPLICATIONS Application includes Microbiology , ARD Labs , Packing , Life science , Medical Storage & R&D activity.
 

HEATING SYSTEM

Heating system: U shaped Nichrome wire heater in SS Sheathing.

Air circulation: FlangeMotor withImpeller andblower. Interior Illumination for working Area.

There will be two nos of blower from 450 Ltr Onwords.

 

 

 

 

DQ,IQ,OQ,PQ

To Help the customers tomeet current regulatory requirements forthe equipment , Hamco can provide DQ,IQ,OQ,PQ Protocols. This Protocol ensures that the system is properly installed, Operating as per our recommendations , Adequately documented and controlled according to regulatory requirements.

Calibration certificate & Test certificate of critical components along with Operation , Maintenance Manual, Wiring diagram , chamber diagram, sensor position diagram provided.

Operation Training of the instruments along with software installation and software

Training is part of our Operation Qualification.

VALIDATION & CALIBRATION Calibration of chamber is provided with Treacability against standard used.

Hamco will give one cycle 24 hours Validation report offsite . ( Validation is putting temp & RH sensors at different position in the chamber confirming Uniformity at different places in the chamber. )

TEMPERATURE SENSOR AGrade RTD ( PT 100 Sensor) . Temperature resolution 0.1C .

 

 

 

 

 

DISPLAY

HMI Display is 7 Color Touch Screen Display of Delta Or Other reputed Make with display shows

set & Process Value of Temp

Active Alarm List

Status of Component.

Component running Hour display

Password Protected Display

Alarm List & its History.

Graph Display of Temperature.

Limited Door Access through Password Protection.

Auto changeover of Sensor on failure.

DOOR ACCESS SYSTEM

(DS)

Limited Door access through Password protection. The person with Authorised password can only open the door with reason . In software we get thedoor access

report whichshowswhoopen thedoor, when & for what reason with time stamp.

AUTO SENSOR CHANGEOVER (SAFETY-1) In the case of Master Temp sensor Fails , it switch over to Standby Temp Sensor.
TEMPERATURE SCANNER (SCN 4) 4 Temp including one Master Sensor Placed at different Position in the chamber to Study the distribution of conditions in the different locations of the chambers and thus helps in Auto- Validate the chamber and helps in easy detection of Drift in the

sensor

EXTERNAL HOOTER (EX.HTR) External Hooter which horns on active alarm works as alarm transmission system
MOBILE ALERT (GSM) You will get SMS Alert on critical utility paramenters like Temp High & Low Alarm,

Humidity High & Low Alarm, Low water Level, Power Fail etc.

EMAIL ALERT

(EMAIL)

You will get EMAIL Alert on critical utility paramenters like Temp High & Low

Alarm, Door Open, Thermostate High, Power Fail etc.

 

 

 

CONTROLLING SYSTEM

The controlling system is Fully PLC Based , It consist of HMI Display , PLC Digital & Analogue cards, Relays, Contactors, Overload Trip devices, MCB.

Separate Connection for Main & UPS ( UPS will be on Buyer Scope). This helps in recording the chamber conditions even during Power fail and capture the event of Power Fail & Resume during the conditions. The control panel has below features.

IP 22 Protection.

Data Storage of 10,000 Readings. If we set data interval of 30 Mins , it can save approx. one year data.

Controls Temp using PID Loops.

Captures all the events Like Power OFF / Resume , Door Open/Close , etc.

Operates as Safety Controller by cutting of Heating device on Over temperature , Cutting of Refrigeration device of Low Temperature etc and generates Alarms and transmit the same alarm to Multiple locations.

RS 485 Communication Port for PC Connectivity.

 

 

 

 

 

SAFETY

Temperature Safety :-

When Temperature deviates higher sidethan their prescribed limits , the Heater will be cut off by switchgears & if still go higher than by Thermostate. It generated alarm andhorn the buzzer if it deviates. Once it reach within the range , the heating system reset automatically.

WhenTemperature deviateslower sidethantheir prescribed limits , the Main contactor will be cutoff by switchgears & if still go lower than by Thermostate. It generated alarm and horn the buzzer if it deviates. Once

it reach within the range , the Refrigeration system reset automatically.

 

Human Safety :-

All the component has individual MCB for Overload and trip protection.

The system has provided with proper earthing.

Sensor Safety :-

The system has been provided with Stand by Sensor, So in case Master sensor Fails, it switch over to Safety sensor thus reduces the down time possibility.

 

ALARM GENERATION

Audio & Visual Alarms are generated for :

Temperature Deviation.

Sensor Failure.

Thermostate High

PC COMMUNICATION (PCI-1) PC Interface communication Card with RS 485 convertor (Hardware) for

Temperature control system

 

 

 

 

 

 

 

SOFTWARE 21

It is Web Based Software. Having Below said Features .

All 21 CFR Part 11 Requirements areMet.

GMP Compliant- Tamper Proof Documentation of Measured data.

Can Connect 20 Chambers at a Time.

Authorised Access Only.

Auto Backup Facility.

Password Expiry & Multi Wrong attempt of Password will block the User.

Auto Logout for no activity during Prescribed Time. Unlimited Users , Multiple login levels.

Can Created Data Log Report & Alarm for Temp & RH.

Scanner & Scanner Alarm Report for Temp & RH.

Event Log Report

Door Access Report.

HMI Audit Report Etc

 

 