HOT AND COLD CHAMBER HMI (PLC 7) MODEL TEMP. RANGE -20C TO +100C WITHOUT HUMIDITY, MULTI SETPOINT PROGRAMMABLE

CERTIFICATIONS OUR CHAMBERS ARE CE Certified Chambers. We as company is ISO 9001:2008 Certified Company.
FEATURES Consist of Chamber (Hamco 46 series) + Ethernet Communication card + HMI (PLC 7) + SOFTW-21 (Web based Software Part 11 21 CFR) + DS (Door Access System) + SAFETY-1
RANGE/HUMIDITY/ACCURACY/UNIFORMITY Temperature Range: -20C to 100C, Temperature Accuracy: +/- 1C, Temperature Uniformity: +/- 2C.
No RH.
RAMP RATE This chamber has No Ramp Up Time facility.
Usually the ramp will be 0.3 to 0.7C/minute.
Optimum Heat load considered.
Hamco Suggest: “The buyer is required to provide heat load specifications prior to placing order to ensure accurate system design and performance.”
CONSTRUCTION GMP MODEL Double walled, PUF Insulation. Inner Body: SS 304 Mirror Finished, Outside: SS 304 Matt Finished. Double door: Inner viewing door + outer insulated door with lock. 50mm Validation Port with silicon rubber seal. 3″ CFC Free Puff Insulation. Stainless steel Trays with adjustable height. Castor wheels with front locks.
LIGHTING Fluorescent Lights automatically On when door opens.
CHAMBER PRINCIPLE Air circulation by Motor & Fan recirculates air through heater and evaporator. Reduces power consumption by 50%.
ICH GUIDELINES Suitable for Stability Studies at:
-20C -10C 0C 25C 50C 75C 100C
Any temperature within range.
Other cycles must be specified before order.
PROGRAMMABLE FEATURES Multi-set point programmable (temperature, ramp time, soak time)
5 programs
32 steps/program
999 repeat cycles
Hamco suggest: “Provide detailed programmable cycle specifications prior to order.”
APPLICATIONS Electronics & Semiconductor, Aerospace, Pharmaceuticals, Automobile, Medical Storage & R&D.
HEATING & COOLING SYSTEM Heating: U-shaped Nichrome wire heater (SS sheathing)
Cooling: Emerson compressor (R134a refrigerant, CFC-free)
1 compressor
Air circulation: Flange Motor with Impeller/blower
Two blowers (from 450L onwards)
DQ,IQ,OQ,PQ Hamco provides protocols for installation/operation verification. Includes calibration/test certificates, manuals, diagrams, and training.
VALIDATION & CALIBRATION Traceable calibration. 24-hour offsite validation report (temperature uniformity test).
TEMPERATURE SENSOR A Grade RTD (PT100 Sensor). Resolution: 0.1C.
DISPLAY 7″ Delta Color Touch Screen HMI shows:
Set/process values
Alarms & history
Component status
Password protection
Live graphs
Programmable outputs
DOOR ACCESS SYSTEM (DS) Password-protected access with audit trail (user, time, reason).
MOBILE ALERT (GSM) Not available (Premium model only)
CONTROLLING SYSTEM PLC-based with 7″ HMI. 10,000 data storage. Ethernet connectivity. IP22 protection.
BIOMETRIC DOOR ACCESS Not available
SAFETY Temperature: Thermostat heater cutoff
Human: Individual MCB protection + earthing
ALARM GENERATION Audio/visual alarms for:
Temperature deviation
Sensor/component failure
Power failure
Door open
Thermostat high
PC COMMUNICATION Ethernet communication card
SOFTWARE 21 Web-based 21 CFR Part 11 compliant software:
Tamper-proof documentation
20-chamber support
Role-based access
Auto-backup/auto-logout
Data/alarm reports

Temp. Range: -20 to + 100 deg C

MODEL NO. CAPACITY DIMENSION IN CMS SHELVES WT IN KG POWER
Standard GMP Cu. Ft Litres Chamber Over All Unit Packed in wooden case VAC Hz Watts
HAMCO46S -3(-20) HAMCO46 G-3(-20) 3 90 45x45x45 60x95x110 2 110 165 230,50 1500
HAMCO46 S-4(-20) HAMCO46 G-4(-20) 4 120 45x45x60 60x95x125 2 150 225 230,50 2000
HAMCO46 S-8(-20) HAMCO46 G-8(-20) 8 200 60x60x60 80x110x125 2 220 300 230,50 2500
HAMCO46S-12(-20) HAMCO46 G-12(-20) 12 324 60x60x90 80x110x155 3 230 320 230,50 2500
HAMCO46S-16(-20) HAMCO46 G-16(-20) 16 450 60x60x125 80x130x175 4 340 430 230,50 4000
HAMCO46S-21(-20) HAMCO46 G-21(-20) 21 600 60x80x125 80x130x195 5 365 465 230,50 4000
HAMCO46S-28(-20) HAMCO46 G-28(-20) 28 800 80x80x125 100x130x210 5 390 500 230,50 4500
HAMCO46S-34(-20) HAMCO46 G-34(-20) 34 1000 80x80x155 100x130x220 5 430 525 230,50 4500
Get In Touch